Maternal Infant Care




Can’t find the answer to your question here? Contact Us.

  • Giraffe Omnibed and Incubator CS1 - Replace clock battery/contact service message

    Replace the SVC Kit Battery CR2032 PN 2089821-001
    after replacing the clock battery, you must enter the service mode and enter the correct time and date.

  • Exchange blender for the Resuscitation Units

    Exchange blender M1091607-X needs to be ordered directly from depot repair center
    800-437-1171 use the Depot repair prompt

  • Giraffe Shuttle Batteries need to be replaced

    Order battery kit PN 5697355 ((Set of 4)
    After the new batteries are installed the reset button on the UPM will need to be pressed to clear the battery fail alarm.

  • Omnibed or Incubator “Fan always on high speed“ message

    Replace the Heat sink sensor 2081533-001

  • Ordering replacement fan motor PN 5513688

    The new fan motor will have new mounting screws and bushings. Discard ALL the old hardware.

  • System error 85/ Stuck in GE Logo screen/ canopy stuck in the open position:

    Disconnect scale connection to the interface panel and restart the bed

  • Ordering the New updated observation light transformer PN 5514157

    Also need to order harness PN 5514159.
    Recommended to also order PNs 2076390-001, M1129877, M1129879

  • System error 9 error code

    Recommend replacing Ribbon cable and wire harness PN 6600-1411-601 & 6600-1558-700
    Also verify Power board fuses F7 & F8. Fuse kit M1118329
    If it still fails, Contact TS could be Control board or Heat engine

  • Poor print quality on CORO 170 and 250CX Series monitors

    Recommend Cleaning Thermal Print head before replacing the print head FRU
    CORO 259CX PN 2025177-056, CORO 170 PN 2021595-001

  • Novii: The Pod will not charge, or the interface intermittently reboots

    Replace them by calling Depot repair and get an Advanced Exchange unit.
    800-437-1171 use the Depot repair prompt

  • Humidifier calibration instructions say to make an adjustment on the control board and can’t find the adjustment pot from the instructions

    The instructions were originally written for the CarePlus incubators and the control board adjustments mentioned in them do not apply to the Giraffe Incubator or Omnibed.

  • The canopy will not go up or down on an Omnibed

    First thing to check is for a shorten canopy hand switch by looking in the Service page to see the different canopy switches and their status.

Terms & Conditions




1. Disclaimer. GE HealthCare “Learn” products, materials, videos, training, and FAQs (collectively, “Learn Material”) do not guarantee that: (i) Customer or its trainees/personnel are fully trained on the means to service, maintain or troubleshoot GE Healthcare equipment; or (ii) the Learn Material will satisfy any licensure or accreditation. Customer must ensure its trainees/personnel are fully qualified to perform the services, maintenance and troubleshooting carried out by them. 2. Intellectual Property and License. GE HealthCare or its licensors own all rights to the intellectual property in the Learn Material. Customer may provide GE HealthCare with feedback related to the Learn Material and GE HealthCare may use it in an unrestricted manner. GE HealthCare grants Customer a non-exclusive, non-transferable, non-sublicensable, revocable license to use the applicable Learn Material for Customer’s internal business purposes only in operating and/or servicing medical equipment. Customer cannot: (i) display or make available the Learn Material to any other entity; (ii) make copies of the Learn Material (without GE HealthCare’s prior written consent); (iii) transfer the Learn Material outside the United States or Customer’s network; (iv) modify, translate or create derivative works based on the Learn Material; or (v) modify markings, labels or notices of proprietary rights of the Learn Material. 3. Confidentiality. Information that is disclosed or provided by GE HealthCare to Customer or Customer’s trainees/personnel through the Learn Material may be confidential to GE Healthcare, and will be identified as such. Customer and its trainees/personnel will treat such information as confidential, meaning they will not disclose the information to third parties unless required by law. 4. Warranty. LEARN MATERIAL IS PROVIDED “AS IS”, WITHOUT WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, WARRANTIES OF PERFORMANCE, NON-INFRINGEMENT, MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. 5. Additional GE HealthCare Policies and Terms. Learn Material may be subject to additional GE HealthCare policies, registrations, and terms and conditions, which are available upon request.